Operation
Memocheck Sim CYP03D
32
Hauser
Path: Menu/Advanc. function
Function
Configuration options
(factory settings in bold)
Error simulation
Cond c
• Polarizat. warn.
• No Cond. display
• Temp. sens. defect
• Sensor supply
Cond c 4-pol
• Polarizat. warn.
• Cond. val. invalid
• Temp. sens. defect
• Broken connector
• Meas. value inval.
• Resist. maximum
Cond i
• Cond. sens. defect
• Cond. val. invalid
• Temp. sens. defect
• Temp. value inval.
• Ind. curr. too high
• Ind. curr. too low
Once you have selected an error using ,
the error is output at the transmitter. To
delete the error again, remove the check
mark again using .
Ramp
Ramp
(continuous)
Ramp (gradual)
Sim. main value
Starting
value
Stop value
Duration
Number of
steps
Time per step
Conductivity
0.001 µS/cm
to 2000 S/cm
0.001 µS/cm
0.001 µS/cm
to 2000 S/cm
2000 S/cm
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
Temperature
-40.0 to
150.0 ˚C
-40.0 °C
-40.0 to
150.0 ˚C
150.0 °C
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
Resistance
0.001 mΩ to
1.0 GΩ
0.001 mΩ
0.001 mΩ to
1.0 GΩ
1.0 GΩ
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
Raw value temp.
-40.0 to
150.0 ˚C
-40.0 °C
-40.0 to
150.0 ˚C
150.0 °C
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s