Memocheck Sim CYP03D
Operation
Hauser
29
Path: Menu/Advanc. function
Function
Configuration options
(factory settings in bold)
Error simulation
Glass SCS failure
Ref. SCS failure
Temp. sens. defect
Glass SCS warning
Ref. SCS warning
Counter spillover
Meas. value inval.
Once you have selected an error using
, the error is output at the
transmitter. To delete the error again,
remove the check mark again using .
Ramp
Ramp
(continuous)
Ramp (gradual)
Sim. main value
Starting
value
Stop value
Duration
Number of
steps
Time per
step
pH value
pH value
-2.0 to 16.0
pH
-2.00 pH
-2.0 to 16.0
pH
16.0 pH
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
Raw v. C1-C2 (pH)
-750.0 to
750.0 mV
-750.0 mV
-750.0 to
750.0 mV
750.0 mV
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
SCS resistance
0.001 MΩ to
1.000 TΩ
0.001 MΩ
0.001 MΩ to
1.000 TΩ
1.000 TΩ
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
Refer. SCS resist.
0.01 to
60.0 kΩ
0.01 kΩ
-2.0 to 16.0
pH
16.0 pH
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
ORP potential
ORP potential
-2.0 to 2.0 V
-2.0 V
-2.0 to 2.0 V
2.0 V
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
ORP [%]
0.0 to
100.0 %
0.0 %
0.0 to
100.0 %
100.0 %
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
Raw val. -C2 (ORP)
-2.0 to 2.0 V
-2.0 V
-2.0 to 2.0 V
2.0 V
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
Raw value C1
-3.0 to 3.0 V
-3.0 V
-3.0 to 3.0 V
3.0 V
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
rH value
rH value
-40.0 to
50.0 rH
-40.0 rH
-40.0 to
50.0 rH
50.0 rH
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s