Operation
Memocheck Sim CYP03D
24
Hauser
8.2.1
Calibration value
The calibration value of the simulated sensor is the value to which the sensor adjustment
refers.
If you change a calibration value, there will be a brief interruption to communication between
the Memocheck Sim CYP03D and the transmitter to allow the transmitter to accept the
calibration settings.
An unfavorable configuration may result in measured values outside the specified
measuring range. This can lead to fault states in the transmitter. More information on
calibration can be found in the Operating Instructions for your transmitter.
8.2.2
Error simulation
You receive a list of possible errors which you can combine with one another.
The error categories in accordance with NAMUR (F, M, C, S) are treated as follows:
• The errors declared as "F" in the transmitter are indicated by a flashing display.
• Errors in other categories are displayed in the transmitter's diagnostic list.
8.2.3
Ramps
Ramp (continuous)
: You define the start and stop value as well as the duration of a ramp
(
Starting value
,
Stop value
,
Duration
).
Ramp (gradual)
: Instead of a total time, you define the number and the duration of the
increments (
Number of steps
,
Time per step
).
Ramp (continuous)
The ramp rises continuously without any
jumps.
Ramp (gradual)
The ramp is gradual.
t
Value
Max.
Min.
A0017397-EN
Value
Max.
Min.
t
A0017398-EN
• The ramp repeats itself automatically until you stop it.
• The ramp is a simple linear function y = mx + n. The linear element "n" is normally equal to
the zero point. The factor "m" is the slope of the line.
• You will obtain a negative slope for the ramp if you set the starting value higher than the
stop value.