Memocheck Sim CYP03D
Operation
Hauser
31
Path: Menu/Advanc. function
Function
Configuration options
(factory settings in bold)
Error simulation
Temp. sens. defect
Sensor supply
Once you have selected an error using ,
the error is output at the transmitter. To
delete the error again, remove the check
mark again using .
Ramp
Ramp
(continuous)
Ramp (gradual)
Sim. main value
Starting
value
Stop value
Duration
Number of
steps
Time per step
ORP potential
-2.0 to 2.0 V
-2.0 V
-2.0 to 2.0 V
2.0 V
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
ORP [%]
0.0 to
100.0 %
0.0 %
0.0 to
100.0 %
100.0 %
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
Temperature
-40.0 to
150.0 ˚C
-40.0 °C
-40.0 to
150.0 ˚C
150.0 °C
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
Raw value
-2.0 to 2.0 V
-2.0 V
-2.0 to 2.0 V
2.0 V
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
Raw value temp.
-40.0 to
150.0 ˚C
-40.0 °C
-40.0 to
150.0 ˚C
150.0 °C
10 to 6000 s
60 s
1 to 200
10
0.5 to 600 s
1 s
8.2.8
Extended functions: Cond c, Cond c 4-pol and Cond i
Path: Menu/Advanc. function
Function
Configuration options
(factory settings in bold)
Calibration value
Conductive
Toroidal
Cell constant
0.001 E-03 to
10.0 cm
-1
10 E-03 cm
-1
0.001 E-03 to
10.0 cm
-1
2.9 cm
-1
Reference value
0.001 mS/cm
to 1.0 S/cm
0.005 mS/cm
0.001 mS/cm
to 1.0 S/cm
100.0 mS/cm
Reference temp.
0.0 to 60.0 ˚C
25.58 °C
0.0 to 60.0 ˚C
25.0 °C
Temp. offset
-10.0 to 10.0 ˚C
0.0 ˚C
Temp. gradient
-3.0 to 3.0
1.0