LT5
Setting
Hauser
69
6.2
Counter display
Counter display procedure
• The scale plate can rotate freely (1 mm (0.04 in) intervals).
• The counter drum changes the number of the No. 1 drum by one digit when the scale
plate makes a full rotation (100 mm (3.94 in)).
1. Remove the indicator cover.
2. Loosen the thread at the center of the scale plate.
3. Rotate the scale plate and align the numerical value of the counter to the first three
digits of the liquid level.
4. Align the scale plate to the pointer, which corresponds to the last two digits of the
liquid level, then tighten the scale plate thread.
If the pointer is indicating a range of between 97 and 03 on the scale plate, the
counter drum will not change instantly. As it changes gradually in unison with
the rotation of the scale plate, the counter will display half values. Thus, in order
to rule out incorrect readings, the counter window and parts of the scale plate are
color coded.
This completes the counter display process.
m
90
95
00
05
10
1
2
3
4
6
7
8
5
A0041232
57
Counter display
1
Counter (yellow)
2
Counter (black)
3
Pointer (red)
4
Thread
5
Range (yellow)
6
No. 1 drum
7
No. 2 drum
8
No. 3 drum
When the pointer (red) is pointing towards the yellow section, the counter will read
the value on the yellow side, and when it is pointing towards the black section, it will
read the value on the black side.
(Example)
• Yellow: 14
000 mm (551.18 in)
• Black: 13
999 mm (551.14 in)