Installation
Liquiline Control CDC90
28
Hauser
Connecting the rinse water
When connecting the water, please pay attention to the following:
• The rinse water pipe is to be provided by the customer.
• The water pressure must be 3-6 bar (43.5-87 psi).
Pay attention to the quality of the rinse water. Particles larger than 100 μm should be
filtered using a water filter.
First measuring point
3
2
4
5
1
6
A0032653
16
Rinsing block with one assembly
1. Rinse the pipe thoroughly.
2. Connect the rinse water (6) to the water connection (5) of the rinsing block (4).
3. Connect the rinse chamber connection (3) on the rinsing block to the rinse
connection (2) of the assembly (1).
Second measuring point
The rinse pipe of the second measuring point goes via the changeover valve.
3
2
4
5
1
6
7
A0033443
17
Rinsing block with two assemblies (first and second measuring point)
1. Rinse the pipe thoroughly.
2. Connect the rinse water (7) to the water connection (6) of the rinsing block.
3. Connect the rinse chamber connection (4) on the rinsing block (5) to the rinse
connection (3) of the changeover valve (2).
4. Connect the rinse connections to the assemblies (1).
Shorten the multihoses
The hoses in the multihose must be altered depending on the distance.
1. Unscrew the M3 multihose from the rinsing block.
Содержание Liquiline Control CDC90
Страница 2: ......
Страница 117: ...Liquiline Control CDC90 Index Endress Hauser 117 Warnings 5 Water pipe 27 Web browser 59 Web server 54 111 Weight 113 ...