Liquiline Control CDC90
Technical data
Hauser
113
Electromagnetic
compatibility
Interference emission and interference immunity as per EN 61326-1:2013, Class A for
Industry
Electrical safety
IEC 61010-1, Class I equipment
Low voltage: overvoltage category II
Environment < 2000 m (< 6562 ft) above MSL
Max. altitude above MSL
< 2000 m (< 6562 ft) above MSL
Pollution degree
The product is suitable for pollution degree 2.
15.11 Mechanical construction
Dimensions
Weight
Device
Weight
Complete device on mounting plate
Approx. 52 kg (114.64 lbs)
CDC90 control unit
Approx. 2.1 kg (4.63 lbs) depending on the version
Pneumatic control unit painted
7.5 kg (16.53 lbs) (empty)
Pump canister unit
Approx. 1.5 kg (3.30 lbs)
Mounting plate (Trespa)
Approx. 10 kg (22 lbs)
Canister shelf
Approx. 3.2 kg (7.05 lbs)
SD card
Max. 5 g (0.17 oz)
Materials
Device
Material
CDC90 control unit
Module housing
PC (polycarbonate)
Soft keys
TPE (thermoplastic elastomers)
LED
POM
Cable mounting rail
Stainless steel 1.4301 (AISI 304)
Display glass
Plastic capacitive touchscreen
Cable glands
PA (polyamide) V0 as per UL94
M12 cable glands
PA (polyamide)
Housing seals
EPDM
Cable gland O-ring
EPDM
Pneumatic control unit
Housing
Stainless steel 1.4301 (AISI 304), painted steel
Housing seals
EPDM (ethylene propylene diene rubber)
Cable glands
PA (polyamide) V0 as per UL94
Housing seals
EPDM