Liquiline Control CDC90
Diagnostics and troubleshooting
Hauser
93
Item
Error message
Troubleshooting
406
Invalid data entry in calibration
settings.
‣
Check buffer values and limit values of
calibration results.
406
Invalid data entry
410
Unstable measured values.
‣
Ensure that the process is stable. Repeat the
calibration. Replace the sensor.
410
Measured value is unstable
411
Calibration calculations are invalid.
1.
Set the correct references (buffer).
2.
Repeat the calibration.
411
Invalid calibration result
‣
412
Calibration settings have not been
made.
‣
Repeat the calibration. Set the correct
references (buffer).
413
An error has occurred during
cleaning.
‣
Ensure that the actuators (pump, float
switches) are connected.
414
An error has occurred during buffer
filling.
1.
Check buffer level.
2.
Check float switches.
415
Sensor - adjustment error
415
Error during sensor adjustment
416
Unexpected command sent.
‣
Please correct the program steps for
calibration.
500
Warning limit for piston
movements exceeded.
‣
Carry out maintenance on the assembly.
501
Alarm limit for piston movements
exceeded.
‣
Carry out maintenance on the assembly.
502
Assembly position not defined.
‣
Check piston.
600600
Error message from the relay
module.
‣
Replace the relay module.
601
Level error
601
No float switch signal.
1.
Fill canister.
2.
Replace float switch.
602
No float switch signal.
1.
Fill canister.
2.
Replace float switch.
603
No float switch signal.
1.
Fill canister.
2.
Replace float switch.
604
Limit for residual volume of
canister liquid reached.
‣
Fill canister.
605
Limit for residual volume of
canister liquid reached.
‣
Fill canister.
606
Limit for residual volume of
canister liquid reached.
‣
Fill canister.
607
Warning limit for delivery volume
of pump 1 reached.
‣
Replace the pump.
608
Alarm limit for delivery volume of
pump 1 reached.
‣
Replace the pump.
609
Warning limit for delivery volume
of pump 2 reached.
‣
Replace the pump.
610
Alarm limit for delivery volume of
pump 2 reached.
‣
Replace the pump.
611
Warning limit for delivery volume
of pump 3 reached.
‣
Replace the pump.