Liquiline Control CDC90
Commissioning
Hauser
63
To return to the start screen, go to
Menu
/
System
/
Meas. point overview
.
9.3.4
Service operating mode
During commissioning, the mode must be set to
Service
.
1. In the menu, go to
Menu
/
Setup
/
Settings
.
2. For
Operating mode
switch to
Service
.
In the main
System
menu, a range of activities can now be performed.
9.3.5
Starting commissioning program
‣
Press and hold soft keys 1 and 3 at the same time for two seconds.
The LED lights up green once the program starts.
The pumps are now activated one after the other and the hoses are filled with
liquids for approx. five seconds.
The program can be stopped by means of a short click on the soft key.
9.3.6
Configuring canister volume
‣
In the menu, go to
Menu
/
Setup
/
System Settings
/
Canister
1. Use
Next
until you reach the point where you enter the canister volume.
2. Under
Initial volume can. 1...3
, enter the volume in liters.
9.3.7
Configuring device
Change the settings for data logging with this menu item, such as the operating mode, the
measuring principle or the type of electrode.
The system is preconfigured in the factory with the following values:
Controllers and actuators
Number
Low power sensor interface
2
High power sensor interface
1
Inputs and outputs
DI
4...16
DO
4...16
AO
2...6
AI
2
Relay
2 ... 4
Actuators
Pilot valves
4...16
Process valves
2...6
Rinsing block
2...6 connections