Operation
Liquiline Control CDC90
72
Hauser
Path:Menu/Setup/Program step
Function
Options
Info
Settings Pilot Valve
Selection
• Used pilot valve
• Assemb. control
• Assembly meas. pos.
• Assembly service pos.
• Water
• Air
• Pump 1 ...3
• Distributor ...16
Factory setting
Depending on program
step
Default setting:
Used pilot valve
Shuts down the program step
Assemb. control
Regulates the pressure in the assembly
Additional options:
Assembly meas. pos.
Assembly is in measuring position
Assembly service pos.
Assembly is in service position.
Water
Rinse water on
Air
Air on
Pump 1
Cleaner
Pump 2
Buffer 1
Pump 3
Buffer 2
Distributor..16
Additional valves
Example: program step 9, cleaner
CPA8xx assembly
Uses pilot valve: Yes
Valve 1: ON
Valve 2: _
Valve 3: OFF
Valve 4: OFF
Valve 5: OFF
Valve 6: OFF
Valve 7: ON
Valve 8: _
Digital inputs
Selection
• Used digital input
•
Assembly
Service pos.
•
Assembly
Meas. Pos.
• DI 3...4
• DI 5...16
Monitor digital inputs 1 to 16 in the pneumatic
control unit/remote IO. The first four inputs are
reserved:
Used digital input
Additional valves
AssemblyService pos.
Response as to whether the assembly of the 1st
measuring point is in service position.
AssemblyMeas. Pos.
Response as to whether the assembly of the 1st
measuring point is in measuring position.
DI 3
Response as to whether the assembly of the 2nd
measuring point is in measuring position.
DI 4
Response as to whether the assembly of the 2nd
measuring point is in measuring position.
Additional options:
DI 5...16
Other digital inputs for configuration
Digital outputs
Selection
DO 1 ...16
In order to be able to assign digital outputs, the
checkbox beside
Used digital input
must be
selected.