8.2.2
Assign soft keys
The soft keys are assigned the following functions:
Key combination
(press
simultaneously)
Function
1 and 2
Switches configuration and visualization of measuring point to 1
2 and 3
Switches configuration and visualization of measuring point to 2
3 and 4
Switches "Service assembly"
1 and 3
Starts elimination of air from liquid hoses
You can assign programs to the soft keys of the CDC90 control unit for the purpose of
moving the measuring points manually.
Proceed as follows:
Select measuring point
1. Go to
Menu
/
Setup
/
System Settings
/
Remote control
2. Use
Next
until you reach the menu where you can configure the soft keys.
3. Select the measuring point in which the program is to run.
4. For
Softkey 1...4
select the programs which you would like to save to the keys.
First configure the soft keys for all measuring points.
Assign soft keys
‣
For
Softkey 1...4
select the programs which you would like to save to the keys.
Manual operation of the soft keys can be applied to only one measuring point. Always
switch to the relevant measuring point.
Starting the program via the soft keys
‣
Press the soft key of the program you wish to start for approx. 2 seconds.
The selected program is displayed on the left-hand side of the start screen. The LED
lights up green once the program starts.
Operation switches automatically to the manual mode. After the program ends, you
must change the mode manually.
The program can be stopped by briefly clicking the soft key or with the key
combination 3 and 4 (Assembly Service).