Operation
Liquiline Control CDC90
66
Hauser
10.1.2 Measuring point
1
2
4
5
6
7
8
9
10
11
3
A0033716
Position
Function
1
Program ID
2
Program step - remaining time
3
Measured value and temperature display
4
Rinsing block valve
5
Pilot valves, lit green during operation
6
Compressed air
7
Canister indicator, indicating residual volume in liters
8
Pump volume delivered in liters
9
Pump, lit green during operation
10
Program step name
11
Program name
Path:Menu/System/Meas. point overview/Meas. Point
Function
Info
Meas. Point
Pump status (green ON / white OFF)
Status of canister level (empty x or not empty)
Relative residual volume of canister, pilot valve status (green on / white off)
process valve status (on/off)
Primary values (e.g. pH temperature)
Assembly position (service/measure position)
Assembly status (defined position/undefined position x); more accurate
analysis depends on the assembly (configure under:
Menu/SETUP/System
settings/Assembly
10.2 Settings
Possibility to configure programs, program steps and calibration functions.
The settings can be modified only the system administrator.