Technical data
Liquiline Control CDC90
108
Hauser
15
Technical data
15.1 Input
Measured values
→ Documentation of the connected sensor
Measuring ranges
→ Documentation of the connected sensor
Types of input
• Digital sensor inputs for sensors with Memosens protocol
• Digital inputs (optional)
Input signal
Depending on version:
• Max. 2 x binary sensor signal
• 2 x 0/4 to 20 mA (optional), passive, potentially isolated from one another and from the
sensor inputs
• 0 to 30 V DC
15.2 Digital inputs, passive
Electrical specification
• drawing power (passive)
• Galvanically isolated
Span
• High: 11 to 30 V DC
• Low: 0 to 5 V DC
Nominal input current
max. 8 mA
PFM function
Minimum pulse width: 500 µs (1 kHz)
Test voltage
500 V
Cable specification
Max. 2.5 mm
2
(14 AWG)
15.3 Current input, passive
Span
> 0 to 20 mA
Signal characteristic
Linear
Internal resistance
Non-linear
Test voltage
500 V