multi EA 4000
Transport and storage
145
5.
Detach the protective conductor (5) from the furnace
cartridge (4) and the L-shaped mounting plate (3).
6.
Detach the orange connection cable of the thermocouple
(1) and the green ground conductor (2) from the inside
wall of the basic device.
7.
Detach the electric supply cables (1.2) from the heating
elements. To do so undo the hexagon socket screws at
the braided wires. Pull out the cable with braided wires
from the top by twisting them to and fro slightly.
8.
Detach the two screws (2, 3) attaching the furnace car-
tridge and mounting plate in the analyzer.
9.
Unscrew the temperature sensor (1) from the furnace
cartridge.
10.
Pull the furnace cartridge and mounting plate fully out
of the analyzer and put them safely aside.
11.
Detach the long braided wire (see arrow) connecting the
two heating elements electrically to each other.
12.
To do so release the hexagon socket screws and pull out
the braided wire from the top by twisting it to and fro
slightly.
Содержание Analytik Jena multi EA 4000
Страница 1: ...Operating Manual multi EA 4000 Elementary analyzer C S and Cl solids analysis...
Страница 50: ...Commissioning multi EA 4000 50 Figure 32 multi EA 4000 C S Hose diagram...
Страница 52: ...Commissioning multi EA 4000 52 Figure 33 multi EA 4000 Cl Hose diagram...
Страница 54: ...Commissioning multi EA 4000 54 Figure 34 multi EA C S Cl for C S measurements Hose diagrams...
Страница 55: ...multi EA 4000 Commissioning 55 Figure 35 multi EA 4000 C S Cl for Cl measurements...
Страница 57: ...multi EA 4000 Commissioning 57 Figure 36 Automatic TIC solids module Hose diagrams...
Страница 58: ...Commissioning multi EA 4000 58 Figure 37 multi EA 4000 C S with automatic TIC solids module...
Страница 60: ...Commissioning multi EA 4000 60 Figure 38 multi EA 4000 with man TIC solids module Hose diagrams...
Страница 61: ...multi EA 4000 Commissioning 61 Figure 39 multi EA 4000 C S Cl with man TIC solids module...