multi EA 4000
Fault removal
135
8.4
Equipment fault
Other problems not detected by the system monitoring can also occur. Starting a
measurement is possible. Such errors are usually detected on the basis of implausible
measuring results (analytical problems). Often, they are clearly recognizable in the
equipment technology. If the suggested solutions are not successful, inform Service.
8.4.1
Basic device
Error
furnace does not heat
Cause
Solution
Malfunction in electronics
Heating element broken.
Fuse in the basic device has tripped.
Check electrical plug connection (mains cable, interface
cable).
Inform Service.
Fans on the rear of the device are not running
Cause
Solution
Fuse faulty.
Electronics faulty.
Switch off the analyzer from the mains switch.
Inform Service.
No intake flow at the "pump" flow regulator
Cause
Solution
Pump not switched on.
Check temperature difference.
Wait for the temperature difference between target and
actual temperature to approximate 50 °C.
If the pump does not activate at a temperature differ-
ence between target and actual temperature of less
than 50 °C, inform Service.
Hoses detached.
Check the hose system for kinked or squeezed hoses.
8.4.2
C/S module
multiWin: Flow error
ACTUAL measuring gas flow < 90 l/h o >110 l/h
Cause
Solution
No intake flow or too low
With the intake pump switched on correct intake flow at
the "pump" flow regulator at the basic module.
Check the hose system for kinked or squeezed hoses.
Desiccant too moist and clumped.
Replace desiccant.
multiWin: Flow error
Measuring gas flow A
CTUAL
= 0
Cause
Solution
No intake flow.
Check temperature difference.
Содержание Analytik Jena multi EA 4000
Страница 1: ...Operating Manual multi EA 4000 Elementary analyzer C S and Cl solids analysis...
Страница 50: ...Commissioning multi EA 4000 50 Figure 32 multi EA 4000 C S Hose diagram...
Страница 52: ...Commissioning multi EA 4000 52 Figure 33 multi EA 4000 Cl Hose diagram...
Страница 54: ...Commissioning multi EA 4000 54 Figure 34 multi EA C S Cl for C S measurements Hose diagrams...
Страница 55: ...multi EA 4000 Commissioning 55 Figure 35 multi EA 4000 C S Cl for Cl measurements...
Страница 57: ...multi EA 4000 Commissioning 57 Figure 36 Automatic TIC solids module Hose diagrams...
Страница 58: ...Commissioning multi EA 4000 58 Figure 37 multi EA 4000 C S with automatic TIC solids module...
Страница 60: ...Commissioning multi EA 4000 60 Figure 38 multi EA 4000 with man TIC solids module Hose diagrams...
Страница 61: ...multi EA 4000 Commissioning 61 Figure 39 multi EA 4000 C S Cl with man TIC solids module...