Operation
multi EA 4000
100
Click on in the sequence row to release it for analysis. All sequence rows can
be released by clicking on the
button.
Sequence rows released for analysis have a green fill.
12.
Click on [S
TART MEASUREMENT
].
The analyzer starts processing the analysis sequence. The M
EASUREMENT
–
A
NALYSIS
window opens.
The boat pick-up picks up
boat 1
from the sample rack and places it onto the
ceramic hook.
The ceramic hook moves the boat into the TIC reactor. As soon as the boat has
reached the position under the ceramic cannula, the acid addition starts auto-
matically. During the slow boat feed acid is added to the entire sample.
The measuring gas from the purged TIC is passed via the measuring gas drying
and cleaning of the automatic TIC solids module to the detector in the C/S
module.
Boat 1
is then pulled out of the TIC reactor and placed onto the ejection rail.
The sampler picks up
boat 2
with the same sample from the sample rack and
moves it directly into the hot zone of the furnace. Here it is incinerated in the
oxygen flow. The total carbon content (TC) is determined in this manner.
The difference of the two measured values (TC-TIC) is then generated and out-
put as TOC result in the program.
During extended measurement interruptions the gas supply can be turned down in the
multiWin program to save gas accordingly. However, the gas supply must be ensured
again in good time (approx. 10 min.) before starting the measurement.
Содержание Analytik Jena multi EA 4000
Страница 1: ...Operating Manual multi EA 4000 Elementary analyzer C S and Cl solids analysis...
Страница 50: ...Commissioning multi EA 4000 50 Figure 32 multi EA 4000 C S Hose diagram...
Страница 52: ...Commissioning multi EA 4000 52 Figure 33 multi EA 4000 Cl Hose diagram...
Страница 54: ...Commissioning multi EA 4000 54 Figure 34 multi EA C S Cl for C S measurements Hose diagrams...
Страница 55: ...multi EA 4000 Commissioning 55 Figure 35 multi EA 4000 C S Cl for Cl measurements...
Страница 57: ...multi EA 4000 Commissioning 57 Figure 36 Automatic TIC solids module Hose diagrams...
Страница 58: ...Commissioning multi EA 4000 58 Figure 37 multi EA 4000 C S with automatic TIC solids module...
Страница 60: ...Commissioning multi EA 4000 60 Figure 38 multi EA 4000 with man TIC solids module Hose diagrams...
Страница 61: ...multi EA 4000 Commissioning 61 Figure 39 multi EA 4000 C S Cl with man TIC solids module...