Maintenance and care
multi EA 4000
108
7.2
Operation with C/S module
Maintenance task
Maintenance interval
Inspect ceramic tube for cracks and visible
damage, replace if necessary
Every three months
Inspect and clean dust trap
Monthly (dependent on combustion and
sample matrix)
Inspect the dust filter, replace in case of visible
contamination
Monthly (dependent on combustion and
sample matrix)
Changing the desiccant in the drying tube
As required (blackening or clumping due to
moisture absorption)
Replace the halide trap
when half of the copper wool or the brass
wool is discolored
7.2.1
Inspect and replace ceramic tube
WARNING
Risk of electric short circuit! In the combustion furnace HTS1500 very high electrical
currents occur in spite of a low voltage (max. 14 V). The combustion tube is part of the
cover of live components.
Only switch on the analyzer with the combustion tube inserted! Always switch off the
analyzer from the mains switch at the rear of the device for conversions and mainte-
nance to the combustion tube or combustion furnace.
CAUTION
There is a risk of burns on the combustion furnace! Only remove the combustion tube
when the device is cold or allow the device to cool down sufficiently (best overnight)!
1.
Switch off the basic device from the mains switch, pull
the mains switch out of the outlet and disconnect the
gas supply.
2.
Create at least 50 cm space on the left side of the basic
device. To do so, move the detector module aside or turn
it to the back to expose the opening of the dust trap.
3.
Pull the contact guard off the opening. The dust trap is
now visible.
Maintenance intervals
Removing the ceramic
tube
Содержание Analytik Jena multi EA 4000
Страница 1: ...Operating Manual multi EA 4000 Elementary analyzer C S and Cl solids analysis...
Страница 50: ...Commissioning multi EA 4000 50 Figure 32 multi EA 4000 C S Hose diagram...
Страница 52: ...Commissioning multi EA 4000 52 Figure 33 multi EA 4000 Cl Hose diagram...
Страница 54: ...Commissioning multi EA 4000 54 Figure 34 multi EA C S Cl for C S measurements Hose diagrams...
Страница 55: ...multi EA 4000 Commissioning 55 Figure 35 multi EA 4000 C S Cl for Cl measurements...
Страница 57: ...multi EA 4000 Commissioning 57 Figure 36 Automatic TIC solids module Hose diagrams...
Страница 58: ...Commissioning multi EA 4000 58 Figure 37 multi EA 4000 C S with automatic TIC solids module...
Страница 60: ...Commissioning multi EA 4000 60 Figure 38 multi EA 4000 with man TIC solids module Hose diagrams...
Страница 61: ...multi EA 4000 Commissioning 61 Figure 39 multi EA 4000 C S Cl with man TIC solids module...