multi EA 4000
Function and setup
29
3.4.3
Measuring gas cleaning
In the operating mode with split, the entire measuring gas from the combustion cham-
ber is divided into a defined ratio. Only some of the measuring gas reaches the meas-
uring cell in order to also permit the determination of high chlorine concentrations.
The discarded share of the measuring gas is routed via the halide trap in the Cl module
to remove the HCl from the gas flow and protect the split valve and intake pump
against corrosion.
The gas extracted from the measuring cell is routed via the adsorption tube. Here the
acidic acid vapor escaping from the electrolyte is removed. Active carbon is used as ad-
sorption agent.
1 Adsorption tube
2 Halide trap
Figure 17 Halide trap and adsorption tube in the Cl module
3.4.4
Connections
The green LED at front of the Cl module illuminates after the module has been
switched on.
Dependent on the equipment of the multi EA 4000 with or without C/S module, the
connections for connecting to the C/S module or the basic device and the autosampler
are located at the rear.
The hoses for the supply of measuring gas and the bypass are routed via the channel
in the basic device through the opening in the right-hand device wall into the chlorine
module.
Halide trap
Adsorption tube
Содержание Analytik Jena multi EA 4000
Страница 1: ...Operating Manual multi EA 4000 Elementary analyzer C S and Cl solids analysis...
Страница 50: ...Commissioning multi EA 4000 50 Figure 32 multi EA 4000 C S Hose diagram...
Страница 52: ...Commissioning multi EA 4000 52 Figure 33 multi EA 4000 Cl Hose diagram...
Страница 54: ...Commissioning multi EA 4000 54 Figure 34 multi EA C S Cl for C S measurements Hose diagrams...
Страница 55: ...multi EA 4000 Commissioning 55 Figure 35 multi EA 4000 C S Cl for Cl measurements...
Страница 57: ...multi EA 4000 Commissioning 57 Figure 36 Automatic TIC solids module Hose diagrams...
Страница 58: ...Commissioning multi EA 4000 58 Figure 37 multi EA 4000 C S with automatic TIC solids module...
Страница 60: ...Commissioning multi EA 4000 60 Figure 38 multi EA 4000 with man TIC solids module Hose diagrams...
Страница 61: ...multi EA 4000 Commissioning 61 Figure 39 multi EA 4000 C S Cl with man TIC solids module...