Operation
multi EA 4000
98
Populate the autosampler. Use the
and
buttons to move the sample
rack to the desired position.
Click on in the sequence row to release it for analysis. All sequence rows can
be released by clicking on the
button.
Sequence rows released for analysis have a green fill.
12.
Click on [S
TART MEASUREMENT
].
The analyzer starts processing the analysis sequence. The M
EASUREMENT
–
A
NALYSIS
window opens.
The boat pick-up picks up a boat from the sample rack and places it onto the
ceramic hook.
The ceramic hook moves the boat into the TIC reactor. As soon as the boat has
reached the position under the ceramic cannula, the acid addition starts auto-
matically. During the slow boat feed acid is added to the entire sample.
The measuring gas from the purged TIC is passed via the measuring gas drying
and cleaning of the automatic TIC solids module to the detector in the C/S
module. The oxygen supply at the inlet of the TIC reactor prevents measuring
gas from escaping through the open inlet.
6.5.4
TOC determination based on differentiation method
CAUTION
During autosampler operation there is a risk of burns at the hot boat and ceramic
hook! After pulling out of the furnace the boat and hook remain hot for a long time.
CAUTION
Crushing hazard at the autosampler! The autosampler arm, the gripper and the boat
carousel move during activation and initialization and measurement operation. Keep
an adequate distance to avoid having your hands crushed.
When filling the autosampler: Only open the cover when the boat carousel is no longer
moving.
This is the method recommended here.
1.
Fill the acid bottle of the TIC solids module with 40% orthophosphoric acid.
2.
Switch on the basic device multi EA 4000, the C/S module, the automatic TIC solids
module and the FPG 48. Make sure to switch on the basic device both from the
mains switch at the rear and the On/Off switch at the front.
The status lights at the front of the devices illuminate.
Do not switch any Cl module present.
3.
Start the multiWin program.
4.
Change between the various analyzer modules. To do so, first activate the sensors:
Содержание Analytik Jena multi EA 4000
Страница 1: ...Operating Manual multi EA 4000 Elementary analyzer C S and Cl solids analysis...
Страница 50: ...Commissioning multi EA 4000 50 Figure 32 multi EA 4000 C S Hose diagram...
Страница 52: ...Commissioning multi EA 4000 52 Figure 33 multi EA 4000 Cl Hose diagram...
Страница 54: ...Commissioning multi EA 4000 54 Figure 34 multi EA C S Cl for C S measurements Hose diagrams...
Страница 55: ...multi EA 4000 Commissioning 55 Figure 35 multi EA 4000 C S Cl for Cl measurements...
Страница 57: ...multi EA 4000 Commissioning 57 Figure 36 Automatic TIC solids module Hose diagrams...
Страница 58: ...Commissioning multi EA 4000 58 Figure 37 multi EA 4000 C S with automatic TIC solids module...
Страница 60: ...Commissioning multi EA 4000 60 Figure 38 multi EA 4000 with man TIC solids module Hose diagrams...
Страница 61: ...multi EA 4000 Commissioning 61 Figure 39 multi EA 4000 C S Cl with man TIC solids module...