On the left-hand side behind the contact guard is the opening for the dust trap with
the connection for the measuring gas.
Left-hand side with dust trap contact guard
Dust trap with measuring gas connection
Figure 9 Connections on the left-hand side of the multi EA 4000
At the rear of the multi EA 4000 are the mains switch, the gas inlets for the oxygen
and argon supply, the connections for the connection to the C/S or Cl module and the
mains connection cable.
Using the mains switch the multi EA 4000 can be completely disconnected from the
mains, which is important for maintenance and repair.
If the mains switch is switched on, the analyzer is in standby mode. For daily operation
the multi EA 4000 is switched on and off from the On/Off switch at the front (1 in Fig-
ure 7). The analyzer is then still e.g. ventilated when switched off. During extended in-
terruptions of measurements lasting several days, the multi EA 4000 must be discon-
nected from the mains using the mains switch.
NOTICE
!
During operation both switches must be on.
1 Mains switch
2 Mains cable
3 "com1" connection to the C/S module
or Cl module
4 "analyte out" connection to the
C/S module
5 "pump in" connection to the C/S
module or Cl module
6 "waste out" connection –
measuring gas outlet
7 Gas connection "Ar" for argon during
Cl measurements and pyrolysis
8 Gas connection "O
2
" for oxygen
The "service 1,2,3" connections are re-
served for customer service.
The "optional fan power" connection is
not occupied.
Figure 10 Connections at the rear of the multi EA 4000
Connections on the
left-hand side
Connections at the
rear
Содержание Analytik Jena multi EA 4000
Страница 1: ...Operating Manual multi EA 4000 Elementary analyzer C S and Cl solids analysis...
Страница 50: ...Commissioning multi EA 4000 50 Figure 32 multi EA 4000 C S Hose diagram...
Страница 52: ...Commissioning multi EA 4000 52 Figure 33 multi EA 4000 Cl Hose diagram...
Страница 54: ...Commissioning multi EA 4000 54 Figure 34 multi EA C S Cl for C S measurements Hose diagrams...
Страница 55: ...multi EA 4000 Commissioning 55 Figure 35 multi EA 4000 C S Cl for Cl measurements...
Страница 57: ...multi EA 4000 Commissioning 57 Figure 36 Automatic TIC solids module Hose diagrams...
Страница 58: ...Commissioning multi EA 4000 58 Figure 37 multi EA 4000 C S with automatic TIC solids module...
Страница 60: ...Commissioning multi EA 4000 60 Figure 38 multi EA 4000 with man TIC solids module Hose diagrams...
Страница 61: ...multi EA 4000 Commissioning 61 Figure 39 multi EA 4000 C S Cl with man TIC solids module...