Transport and storage
multi EA 4000
146
13.
Successively lift out the heating elements carefully from
the furnace cartridge.
14.
Package the two heating elements and the braided wire
and the furnace cartridge in separate transport contain-
ers.
9.1.3
Transport notes
Observe the safety instructions in section "Safety instructions, transport and " p. 12.
Transport the analyzer very carefully to prevent damage from impact or vibration. For
the transport the heating elements and furnace cartridge must be removed and pack-
aged separately in the transport container. Otherwise, there is a risk of breakage!
The analyzer should be transported in such a way that major temperature fluctuations
are avoided and the formation of condensate is thus prevented in the transport con-
tainer.
9.1.4
Moving the analyzer in the laboratory
CAUTION
Unintentional dropping of the analyzer poses a risk of injury and the analyzer will be
damaged!
Move the analyzer with great care! 2 persons are required to lift and carry the ana-
lyzer!
Observe the following when moving the analyzer within the laboratory:
Insufficiently secured components pose a risk of injury! Before moving the analyzer
remove all loose components, in particular the reagent bottle with phosphoric acid
and the sulfuric acid container.
Disconnect all supply connections and add-on devices from the analyzer. For a
transport within the laboratory the heating elements and furnace cartridge need
not be removed from the analyzer.
To prevent health damage the following must be observed when moving the ana-
lyzer in the laboratory (lifting and carrying):
Содержание Analytik Jena multi EA 4000
Страница 1: ...Operating Manual multi EA 4000 Elementary analyzer C S and Cl solids analysis...
Страница 50: ...Commissioning multi EA 4000 50 Figure 32 multi EA 4000 C S Hose diagram...
Страница 52: ...Commissioning multi EA 4000 52 Figure 33 multi EA 4000 Cl Hose diagram...
Страница 54: ...Commissioning multi EA 4000 54 Figure 34 multi EA C S Cl for C S measurements Hose diagrams...
Страница 55: ...multi EA 4000 Commissioning 55 Figure 35 multi EA 4000 C S Cl for Cl measurements...
Страница 57: ...multi EA 4000 Commissioning 57 Figure 36 Automatic TIC solids module Hose diagrams...
Страница 58: ...Commissioning multi EA 4000 58 Figure 37 multi EA 4000 C S with automatic TIC solids module...
Страница 60: ...Commissioning multi EA 4000 60 Figure 38 multi EA 4000 with man TIC solids module Hose diagrams...
Страница 61: ...multi EA 4000 Commissioning 61 Figure 39 multi EA 4000 C S Cl with man TIC solids module...