Removing the Top Cover
Figure 7-12:
A.
Top cover
B.
M5 x 12 mm button head socket screws (4)
C.
Chassis assembly
D.
M5 flat washers (4)
Procedure
1.
Set the
ON/OFF
switches on the front and back panel to OFF.
2.
Allow the analyzer to cool for at least two hours. Allow it to cool for twelve hours if
replacing the analysis cell.
3.
Disconnect all external cables.
Tag if necessary.
4.
Remove the analyzer from the rack and place on a flat, stable surface that can
accommodate the weight. This step requires two people to lift and transport the
analyzer.
Troubleshooting and diagnostics
Installation, Operations, and Maintenance Manual
71