The status outputs (H) provide fault indications to your process control center. Each digital
output is connected to a normally closed relay, located inside the analyzer, which will open
in response to the detection of a specific fault. The possible causes of a fault indication are:
1.
The sample gas concentration is outside of specification, i.e., the sample gas
concentration has exceeded the measurement range of the instrument.
2.
The analyzer is out of specification or has developed a fault.
WARNING!
HIGH VOLTAGE
Voltages up to 250 Vac, 50 Hz may be present on the digital output terminals.
External circuits should be installed in accordance with national wiring regulations.
Failure to obey the wiring regulations may result in serious injury to personnel.
WARNING!
ELECTRIC SHOCK
The analyzer passed electromagnetic compatibility (EMC) tests based on all electrical cables
and harnesses attached to the instrument having a length of 3 m (9.8 ft.) Attaching cables and
wiring harnesses longer than 3 m (9.8 ft.) may cause injury to personnel.
2.5
Commissioning the analyzer
Once the analyzer is fully installed as described, it should be commissioned according to
local government regulations and the commissioning plan agreed upon between you and
Emerson.
Install
Installation, Operations, and Maintenance Manual
19
Содержание Rosemount CT5400
Страница 15: ...Abbreviation Description WEEE Waste electrical and electronic equipment μm Micro meter ...
Страница 16: ......
Страница 20: ...Contents iv Rosemount CT5400 ...
Страница 32: ...Plan 12 Rosemount CT5400 ...
Страница 40: ...Install 20 Rosemount CT5400 ...
Страница 44: ...Startup procedure 24 Rosemount CT5400 ...
Страница 54: ...Operating the analyzer 34 Rosemount CT5400 ...
Страница 158: ...Troubleshooting and diagnostics 138 Rosemount CT5400 ...
Страница 164: ...Shutdown procedure 144 Rosemount CT5400 ...
Страница 166: ...Preventative maintenance 146 Rosemount CT5400 ...
Страница 172: ...Engineering Drawings 152 Rosemount CT5400 ...