Procedure
1.
Shut down the analyzer as described in
Section 8.2
and allow it to cool..
2.
Remove the analyzer from the rack and place it on a stable workbench. This step
requires two people to properly lift and transport to the workbench.
3.
Remove and retain the four screws attaching the top cover to the chassis.
4.
Open the locking arms of the ribbon cable connector (B) and disconnect the ribbon
cable (H) from the TEC Board (C).
Replace the TEC Board
Figure 7-22:
A. PCB mounting plate
B. Ribbon cable connector
C. TEC board
D. Spacer
E. M3 flat washer
F. M3 spring washer
G. M3 x 6 mm socket head cap screw
H. Ribbon cable to Motherboard
I. Wiring harness
J. Wiring harness to USB port
K. Motherboard
Troubleshooting and diagnostics
90
Rosemount CT5400