This screen gives a reading of the concentration of the selected gas that is present as
an impurity in the nitrogen calibration gas, as measured by the instrument. In the
example (
Figure 5-4
), the gas being measured is NH
3
(ammonia), and the instrument
has detected a concentration of 0.40 ppm.
11.
If the reading is within tolerance, no further action is required. Press
to end the
zero verification process.
The display controller proceeds to the
Result
screen (
Figure 5-5
).
Result Screen
Figure 5-5:
12.
If the reading is outside tolerance, the instrument should be zero calibrated.
Refer to
Section 6.3.1
.
13.
In either case, press
to return to the
Calibration / Verification Complete
screen
(
Figure 5-6
).
Calibration / Verification Complete Screen
Figure 5-6:
Verifying Gas Concentrations
38
Rosemount CT5400