7.16
Replacing the motherboard
Complete the following steps to replace the motherboard.
Observe all safety precautions before starting this procedure.
DANGER!
ELECTRIC SHOCK
The analyzer operates using mains voltage that is dangerous to life. Make sure that the power
ON/OFF
switch at the rear of the panel is set to
OFF
and tagged off before removing the top
cover.
The analyzer must be earthed.
Death or personal injury may result if this is not observed.
To gain access to the electrical, pneumatic, and optical components it is necessary to
remove the top cover.
Procedure
1.
Shut down the analyzer as described in
Section 8.2
.
2.
Power down the analyzer and allow it to cool.
3.
Remove the analyzer from the rack and place it on a stable workbench. This step
requires two people to properly lift and transport to the workbench.
4.
Tag or otherwise identify all wiring harnesses before disconnecting.
Troubleshooting and diagnostics
Installation, Operations, and Maintenance Manual
95
Содержание Rosemount CT5400
Страница 15: ...Abbreviation Description WEEE Waste electrical and electronic equipment μm Micro meter ...
Страница 16: ......
Страница 20: ...Contents iv Rosemount CT5400 ...
Страница 32: ...Plan 12 Rosemount CT5400 ...
Страница 40: ...Install 20 Rosemount CT5400 ...
Страница 44: ...Startup procedure 24 Rosemount CT5400 ...
Страница 54: ...Operating the analyzer 34 Rosemount CT5400 ...
Страница 158: ...Troubleshooting and diagnostics 138 Rosemount CT5400 ...
Страница 164: ...Shutdown procedure 144 Rosemount CT5400 ...
Страница 166: ...Preventative maintenance 146 Rosemount CT5400 ...
Страница 172: ...Engineering Drawings 152 Rosemount CT5400 ...