After a few seconds, the
Gas Sensor Main
screen (
Figure 3-2
) appears on the display
controller. If it does not, report the fault to the maintenance organization.
Gas Sensor Main screen
Figure 3-2:
6.
Start up the system for returning the sample gas.
7.
Start up the gas handling system that conditions the sample gas before it is fed into
the analyzer. Allow the analyzer to warm up in accordance with
Table 1-3
before the
sample gas supply is turned on.
Startup procedure
Installation, Operations, and Maintenance Manual
23