1.5
Safety precautions and conditions for safe use
WARNING!
Before installing or performing any maintenance on the analyzer, read and understand the
safety information given in the preliminary information of this manual.
The analyzer described in this manual has been quality control tested and left the
manufacturer in pristine condition. To achieve the correct and safe operation of this
product, it must be transported, installed, operated, and maintained as described by the
manufacturer.
All lasers used within the instrument are Class 1. The emitted laser light is invisible (mid-
infrared) and the pulse duration so short that the unprotected eye will not be damaged.
The nature of the laser beam path and beam width further ensures that it should be
impossible to cause any eye damage. The instrument has warning labels at appropriate
positions in accordance with USA 21 CFR 1040.10.
Conditions for safe use
•
This equipment has flamepaths which differ from those in IEC60079-1/EN 60079-1.
Cascade Technologies, Ltd (
) must be contacted for guidance
when maintaining the flamepaths.
•
The fasteners which secure the cover are non-standard and shall therefore only be
replaced by fasteners supplied by the manufacturer for this purpose. The fasteners
must always be fitted with the washer supplied by the manufacturer.
•
The equipment has non-conductive surfaces which are a potential electrostatic
charging hazard - see instructions for guidance.
•
The process gas flow rate is limited to a maximum of 6 liters per minute.
•
The equipment shall only be used with process gases which are in gas groups B, C or
D (Divisions) or IIB + H 2 (Zones) and must not contain oxygen or any other oxidizer
in concentrations greater than that found in normal air.
CSA Certificate North American conditions
•
The equipment has flameproof joints with dimensions which are other than those
specified in Table 2 of ANSI/UL 60079-1: 6th edition and Table 3 of CSA C22.2
60079-1:16. These flameproof joints are not intended to be repaired, but where
necessary the original manufacturer shall be contacted for guidance and
information on the dimensions of the flameproof joints.
•
The fasteners which secure the cover are non-standard and shall therefore only be
replaced by fasteners supplied by the manufacturer for this purpose. The fasteners
must always be fitted with the washer supplied by the manufacturer.
•
The equipment has non-conductive surfaces which are a potential electrostatic
charging hazard – see the instructions for guidance.
•
The user shall ensure that the flow of process gas is limited to a mazimum flow rate
of 6 liters per minute.
Plan
Installation, Operations, and Maintenance Manual
5
Содержание Rosemount CT5400
Страница 15: ...Abbreviation Description WEEE Waste electrical and electronic equipment μm Micro meter ...
Страница 16: ......
Страница 20: ...Contents iv Rosemount CT5400 ...
Страница 32: ...Plan 12 Rosemount CT5400 ...
Страница 40: ...Install 20 Rosemount CT5400 ...
Страница 44: ...Startup procedure 24 Rosemount CT5400 ...
Страница 54: ...Operating the analyzer 34 Rosemount CT5400 ...
Страница 158: ...Troubleshooting and diagnostics 138 Rosemount CT5400 ...
Страница 164: ...Shutdown procedure 144 Rosemount CT5400 ...
Страница 166: ...Preventative maintenance 146 Rosemount CT5400 ...
Страница 172: ...Engineering Drawings 152 Rosemount CT5400 ...