Begin Calibrate span Screen
Figure 6-14:
11.
Press
to confirm. The Calibrate span screen
Figure 6-15
opens.
Calibrate span Result Screen
Figure 6-15:
12.
Allow the concentrations to stabilize and wait for two minutes after stabilization.
This screen gives a reading of the concentration of the selected gas that is present,
as measured by the instrument. In
Figure 6-15
, the gas being measured is NH
3
(ammonia), and the instrument has detected a concentration of 119.41 ppm (vol).
13.
If the reading is within tolerance, no further action is required. Press
or
to
abort the span calibration process.
14.
If the reading is outside tolerance, press
ENTER
.
The analyzer returns to its factory-set calibration.
15.
Allow a minute to ensure the readings are stable.
16.
If the reading is now within tolerance, no further action is required. Press
or
to abort the span calibration process.
Gas Calibration Procedures
54
Rosemount CT5400
Содержание Rosemount CT5400
Страница 15: ...Abbreviation Description WEEE Waste electrical and electronic equipment μm Micro meter ...
Страница 16: ......
Страница 20: ...Contents iv Rosemount CT5400 ...
Страница 32: ...Plan 12 Rosemount CT5400 ...
Страница 40: ...Install 20 Rosemount CT5400 ...
Страница 44: ...Startup procedure 24 Rosemount CT5400 ...
Страница 54: ...Operating the analyzer 34 Rosemount CT5400 ...
Страница 158: ...Troubleshooting and diagnostics 138 Rosemount CT5400 ...
Страница 164: ...Shutdown procedure 144 Rosemount CT5400 ...
Страница 166: ...Preventative maintenance 146 Rosemount CT5400 ...
Страница 172: ...Engineering Drawings 152 Rosemount CT5400 ...