13.
If the reading is within tolerance, no further action is required. Press
to end the
zero calibration process.
14.
If the reading is outside tolerance, press
.
The analyzer returns to its factory set calibration. Allow a minute to ensure the
readings are stable.
15.
If the reading is now within tolerance, no further action is required. Press
or
to abort the
Zero Calibration
process.
16.
If the calibration of the analyzer remains outside of tolerance, press
to display
the Calibrate screen (
Figure 6-8
) to re-calibrate with a new zero offset value.
Calibrate screen (Confirmation)
Figure 6-8:
17.
To proceed with the automatic re-calibration, press
.
The automatic re-calibration of the instrument's zero point starts, and the offset
shown on the screen (
-0.852 ppm (vol)
in
Figure 6-8
) is applied to the instrument.
Press the UP
or DOWN
button to abort the calibration process.
When the calibration is complete,
Figure 6-9
is displayed.
Gas Calibration Procedures
50
Rosemount CT5400
Содержание Rosemount CT5400
Страница 15: ...Abbreviation Description WEEE Waste electrical and electronic equipment μm Micro meter ...
Страница 16: ......
Страница 20: ...Contents iv Rosemount CT5400 ...
Страница 32: ...Plan 12 Rosemount CT5400 ...
Страница 40: ...Install 20 Rosemount CT5400 ...
Страница 44: ...Startup procedure 24 Rosemount CT5400 ...
Страница 54: ...Operating the analyzer 34 Rosemount CT5400 ...
Страница 158: ...Troubleshooting and diagnostics 138 Rosemount CT5400 ...
Страница 164: ...Shutdown procedure 144 Rosemount CT5400 ...
Страница 166: ...Preventative maintenance 146 Rosemount CT5400 ...
Страница 172: ...Engineering Drawings 152 Rosemount CT5400 ...