WARNING!
BURNS
Do not touch any part of the analyzer. Assume all parts of the analyzer are hot unless it has
been switched off and allowed to cool down. Some parts of the analyzer may be heated to
190 °C (374 °F). Always wear proper protective equipment when handling the analyzer.
•
Switch off the analyzer and allow it to cool for at least two hours before fitting,
removing, or performing any maintenance.
•
The analysis cell is insulated against heat loss. Allow the analyzer to cool for at least
twelve hours before performing any maintenance on, or in the vicinity of, the analysis
cell.
Failure to allow sufficient cooling may cause serious burn injury to personnel. If burns occur,
seek immediate medical treatment.
Replacing a temperature controller
Figure 7-26:
A.
Temperature controller
B.
Temperature controller base
C.
Lever
Troubleshooting and diagnostics
Installation, Operations, and Maintenance Manual