5 - 8
Resuscitation Module Controls, Indicators, and Connectors
The actual display may differ in appearance or configuration.
Name
Description
Suction
gauge
Displays the suction level from 0 mmHg (0 cmH
2
O) to
200 mmHg (107 cmH
2
O) of vacuum.
Suction control
Adjusts the suction level from 0 mmHg (0 cmH
2
O) to 150 mmHg
(80 cmH
2
O).
Suction
On/Off
switch
Turns the suction on and off. (Gas supply must be turned on.)
Patient Outlet
Accepts the primary gas tubing. If the blender option is installed,
it delivers blended gas. Otherwise, it delivers 100% oxygen.
Flow Rate
(LPM) control
Adjusts the patient gas flow from 0 LPM to 15 LPM.
Airway Pressure
port (if equipped) Provides a connection for the airway pressure gauge to the breath-
ing circuit.
Airway Pressure gauge (only works
with Primary Flow)
Displays the airway pressure from -1 kPa (-10 cmH
2
O) to 8 kPa
(80 cmH
2
O).
Patient Outlet 2
Connection for the gas delivery line at the hose barb connection
on the front panel. If the blender option is installed, it delivers
blended gas. Otherwise, it delivers 100% oxygen. Use tubing with
an inner diameter of 0.25" (0.64 cm).
Flow Rate 2
(LPM) control
Adjusts the patient gas flow from 0 LPM to 15 LPM for
Patient
Outlet 2
.
Содержание RW82
Страница 2: ...This page intentionally left blank ...
Страница 4: ...This page intentionally left blank ...
Страница 16: ...PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04 1 8 This page intentionally left blank ...
Страница 118: ...PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04 6 10 Pneumatic System Pneumatic System Block Diagram ...
Страница 124: ...PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04 7 4 This page intentionally left blank ...
Страница 125: ...This page intentionally left blank ...