WARNING:
When raising or lowering the upper post of the variable height warmer, ensure that any attached
cables, tubing, or hoses are not compromised. Patient or personal injury, or equipment damage
could occur.
WARNING:
Risk of equipment damage and risk of injury. Reusing, reprocessing, or sterilizing disposable
items can lead to a failure of the medical device and cause injury to the patient. Do not reuse,
reprocess, or sterilize disposable items. Disposable items were designed, tested, and
manufactured for one-time use only.
CAUTION:
To avoid accidental lifting or crushing of objects before raising or lowering the VHA mattress,
ensure the area around the warmer is clear. Failure to do so could cause equipment damage.
Electrical Precautions
SHOCK HAZARD:
An electrical shock hazard exists within the warmer module and controller. Any substitution of
components within the controller or warmer module may impair the intrinsic safety of the unit.
Qualified personnel must service the unit. Failure to do so could result in patient injury, personal
injury, or equipment damage.
SHOCK HAZARD:
Plug the power cable only into a properly grounded wall receptacle approved for hospital use
and of the correct voltage. Do
not
use extension cords or an AC receptacle box for this device.
Personal injury or equipment damage could occur.
SHOCK HAZARD:
Use only the power cables supplied with the unit. Failure to do so could result in personal injury
or equipment damage.
SHOCK HAZARD:
Do not remove the back cover; an electric shock hazard exists inside. Refer servicing to
qualified personnel. Personal injury or equipment damage could occur.
Explosion Precautions
WARNING:
When performing cleaning and maintenance procedures, confirm that the oxygen supply is
turned off and that the equipment is disconnected from the oxygen supply. A fire and explosion
hazard exists when performing cleaning and maintenance procedures in an oxygen-enriched
environment. Personal injury or equipment damage could occur.
WARNING:
Do not use the equipment in the presence of flammable anesthetic gases or other flammable
materials, such as some types of cleaning fluids. Personal injury or equipment damage could
occur.
Содержание RW82
Страница 2: ...This page intentionally left blank ...
Страница 4: ...This page intentionally left blank ...
Страница 16: ...PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04 1 8 This page intentionally left blank ...
Страница 118: ...PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04 6 10 Pneumatic System Pneumatic System Block Diagram ...
Страница 124: ...PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04 7 4 This page intentionally left blank ...
Страница 125: ...This page intentionally left blank ...