PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
4 - 21
Resuscitation Equipment (Optional)
Gas Source Pressure
1. Ensure that the oxygen and air pipelines are securely attached to the appropriate fittings on the rear of
the unit. Also ensure that the gas supply present is 276 kPa (40 psi) to 517 kPa (75 psi).
2. If using the reserve gas supply from cylinders, perform the following:
a. Ensure that the cylinders are properly secured in the mounting yokes on the rear of the warmer
before opening.
b. Ensure that the cylinder valve located on the top of the cylinder is open.
c. Examine the appropriate cylinder pressure gauges on the front of the Gas Supply Module to
ensure that sufficient reserve gas supply is present.
d. Set the handle of the
Gas Supply
On/Off
switch to the On position.
NOTE:
When both wall and cylinder gas sources are turned on, the system may not always pull gas from the wall
source. It pulls from the source with the highest pressure. Check cylinder gauges to determine gas source.
Blended Gas Supply (Optional)
WARNING:
If either the air or oxygen gas source pressure is reduced, creating a pressure differential of
207 kPa (30 psi), the microblender alarm sounds. This condition can alter the FiO
2
and flow
output from the microblender. Patient injury could occur.
WARNING:
Always operate the system with medical-grade gases. Failure to do so could result in patient
injury or equipment damage.
WARNING:
Use air inlet water filters with the low-flow microblender. Replace them in accordance with the
manufacturer recommendation. Failure to do so could result in patient injury or equipment
damage.
WARNING:
Monitor the concentration of oxygen provided and the partial pressure of oxygen in the patient
blood (PaO
2
). Failure to do so could result in patient injury.
If present, set the precision blender to the desired oxygen percentage using the blender control knob.
NOTE:
For blender to operate, the
Gas Supply
On/Off
switch must be on.
NOTE:
In the US, the blender is standard with the AutoBreath Resuscitation Module.
Содержание RW82
Страница 2: ...This page intentionally left blank ...
Страница 4: ...This page intentionally left blank ...
Страница 16: ...PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04 1 8 This page intentionally left blank ...
Страница 118: ...PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04 6 10 Pneumatic System Pneumatic System Block Diagram ...
Страница 124: ...PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04 7 4 This page intentionally left blank ...
Страница 125: ...This page intentionally left blank ...