4 - 15
Operational Checkout Procedure
Perform the operational checkout procedure before each use and after disassembly for cleaning,
servicing, or maintenance. If the unit does not perform properly, do not use it and refer it to qualified
service personnel.
Controller
SHOCK HAZARD:
If the controller or other parts of the equipment fail to function properly, do not use the warmer.
A potential shock hazard exists within the warmer module. Refer service to qualified personnel.
Failure to do so could result in patient injury or equipment damage.
WARNING:
To prevent injury or damage to the warmer, employ two persons of sufficient strength to control
the warmer during transport. Use the handle when moving the equipment.
WARNING:
For optimum stability, always lower the Variable Height Resuscitaire Radiant Warmer to its
lowest position before transport. Ensure that items placed on the shelves are properly secured.
Failure to do so could result in patient injury, personal injury, or equipment damage.
1. Plug the AC line cord into the power connector on the rear panel.
SHOCK HAZARD:
Plug the power cable only into a properly grounded wall receptacle approved for hospital use
and of the correct voltage. Do
not
use extension cords or an AC receptacle box for this device.
Personal injury or equipment damage could occur.
2. Plug the power cable into an appropriate power source.
3. Switch on the On/Off switch. The selftest function initiates, and the following occurs:
a. The
APGAR Timer
,
Baby Temperature
, and
Set Temperature
digital displays show all eights.
b. All alarm indicators light, except for
Power Fail.
c. All mode indicators light.
d. The
>37 °C indicator
lights.
e. All ten segments of the
Heater Power
indicator light.
f. The
Procedural Silence
indicator lights.
g. The
Keypad Lock
indicator lights.
h. The alarm sounds a high-pitched tone, a low-pitched tone, and then three beeps.
4. When the selftest function completes, the controller begins operating in Pre-Warm Mode.
5. Check Pre-Warm Mode:
a. The
Pre-Warm
indicator is on.
b. The
Heater Power
indicator displays 10 segments (100%) for 3 min.
c. The
Heater Power
indicator reduces to six segments (60%) for 12 min.
Содержание RW82
Страница 2: ...This page intentionally left blank ...
Страница 4: ...This page intentionally left blank ...
Страница 16: ...PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04 1 8 This page intentionally left blank ...
Страница 118: ...PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04 6 10 Pneumatic System Pneumatic System Block Diagram ...
Страница 124: ...PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04 7 4 This page intentionally left blank ...
Страница 125: ...This page intentionally left blank ...