PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
2 - 4
WARNING:
The internal adjustable airway pressure relief is not functional with a self-inflating bag, because
the patient supply is connected to the fresh gas reservoir on the self-inflating bag, rather than
directly to the patient airway. When using a self-inflating bag, always use a suitable external
airway pressure relief valve connected to the patient airway. To obtain maximum flow, turn the
adjustable airway pressure relief knob to the
Max.
position. Failure to do so could result in
patient injury.
WARNING:
Using a breathing circuit other than the one supplied may inadvertently restrict the gas flow
such that the indication of the gas flow and the adjustable airway pressure relief are inaccurate.
Patient injury could occur.
•
Patient breathing and supply circuits
—Use the disposable breathing circuit (MP00310) on the
Resuscitation Module.
NOTE:
VentStar Resuscitaire disposable hose system is designed for use with a Resuscitaire® Radiant Warmer.
Refer to VentStar Instructions for Use.
NOTE:
VentStar Resuscitaire disposable hose system must only be used for one patient and must be used
exclusively in manual mode.
Variable oriface
Connect to mask or ET tube
Corrugated tubing
Connect to
Patient outlet
Содержание RW82
Страница 2: ...This page intentionally left blank ...
Страница 4: ...This page intentionally left blank ...
Страница 16: ...PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04 1 8 This page intentionally left blank ...
Страница 118: ...PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04 6 10 Pneumatic System Pneumatic System Block Diagram ...
Страница 124: ...PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04 7 4 This page intentionally left blank ...
Страница 125: ...This page intentionally left blank ...