PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
4 - 11
Assembly of the Resuscitaire Wall Mounted Radiant Warmer (WMRW82)
WARNING:
The wall mount bracket installation must be appropriately certified that it meets or exceeds the
pull-out force before mounting the Resuscitaire Wall Mounted Radiant Warmer. Failure to do so
can result in an unsafe installation that could cause equipment damage or personal or patient
injury.
NOTE:
Due to space constraints, the Resuscitaire Wall Mounted Radiant Warmer must be fully assembled
(except the wall mount cover plate) before it is mounted on the wall.
1. Remove the back cover (A) from the modified post (B)
by removing the two screws (C).
2. Remove the controller (D) by removing the four
screws (E).
3. Mount the modified post (B) on the wall mount bracket
(F) using the wall mount plates (G), the 1/4 - 20 x 3/4”
cap screws and the lock washers (H) in the front holes,
and the 1/4 - 20 x 7/8” cap screws and the lock washers
(I) in the rear holes.
A
B
C
D
E
B
F
G
H
I