PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
2 - 5
Resuscitation Module with External Pressure Manometer and 15 mm Fitting
•
Patient Outlet (Resuscitation Module with External
Pressure Manometer and 15 mm Fitting only)
—Use
the Patient Outlet to provide continuous gas flow to the
patient. Controls are provided for
Airway Pressure
Relief
(maximum pressure) and
Flow Rate
(0 - 15
LPM). The adjustable
Airway Pressure Relief
control
is always operative when used with the 15-mm tubing
provided. Pressure relief is adjustable and ranges from 0
to 4.9 kPa (0 to 50 cmH
2
O/mbar). The fixed internal
safety valve is always operable and provides redundant
maximum pressure relief at
4.9
kPa (50 cmH
2
O/mbar) ±
20%. This relief allows the patient to inspire room air in
the event of gas supply failure.
NOTE:
A one-way valve is installed at the Patient Outlet
connection. This valve opens when pressure in the hose
delivering the gas to the patient falls below -0.4 kPa (-0.06
psi).
NOTE:
Adjust flow rate to between 5 LPM and 15 LPM.
•
Airway Pressure
—The airway pressure gauge monitors the airway pressure when externally
connected to patient circuits.
•
Blender
—The optional blender supplies blended gas only to the Patient Outlets in the Resuscitation
Module. During operation, the resuscitation module alarms if both oxygen or air pressures are low.
•
Patient Outlet 2
—Patient Outlet 2 supplies gas through the
Flow Rate 2
control to the Patient Outlet
2 connector. If the blender option is installed, it delivers blended gas. Otherwise, it delivers 100%
oxygen. It is intended for oxygen enrichment of a resuscitation bag or for blow-by. The
Flow Rate 2
control adjusts the flow rate from 0 LPM to 15 LPM. An internal pre-set relief valve limits the
Patient
Outlet 2
pressure to 15.7 kPa (160 cmH
2
O/mbar).
NOTE:
Adjust gas supply to desired flow rate between 5 LPM and 15 LPM. If using a nasal cannula, adjust flow
rate between 1 LPM to 15 LPM.
Содержание RW82
Страница 2: ...This page intentionally left blank ...
Страница 4: ...This page intentionally left blank ...
Страница 16: ...PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04 1 8 This page intentionally left blank ...
Страница 118: ...PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04 6 10 Pneumatic System Pneumatic System Block Diagram ...
Страница 124: ...PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04 7 4 This page intentionally left blank ...
Страница 125: ...This page intentionally left blank ...