PROPRIETARY AND CONFIDENTIAL DRAFT 8 Jan 04
3 - 8
WARNING:
When using the Patient Outlet, use the 10-mm tubing offered by the manufacturer, or use infant
resuscitation bags with built-in pressure release during infant resuscitation. Failure to do so
could result in patient injury.
WARNING:
Using a breathing circuit other than the one supplied may inadvertently restrict the gas flow
such that the indication of the gas flow and the adjustable airway pressure relief may be
inaccurate. Patient injury could occur.
WARNING:
Always keep the oxygen and air gas supplies clean and dry. Use water trap-filters in the supply
lines if necessary. Failure to do so could result in patient injury or equipment damage.
WARNING:
Securely fasten all hoses to fittings. Hand-tighten to avoid damaging the fittings. Failure to do so
could result in patient injury or equipment damage.
WARNING:
Require wall gas supplies at 276 kPa (40 psi) to 517 kPa (75 psi). Failure to operate at this
range could result in patient injury or equipment damage.
WARNING:
Do not place flow restrictions in the supply line. Patient injury or equipment damage could
occur.
WARNING:
If a compressor is used as the air source, take steps to filter, cool, and dehumidify the
compressed room air before introducing it into the gas supply or primary supply module. Failure
to do so could result in patient injury or equipment damage.
WARNING:
Do not place a humidifier in the supply line. Patient injury or equipment damage could occur.
WARNING:
Gas going to the patient must not be super-saturated, as evidenced by excessive condensation
in the tubing. Patient injury or equipment damage could occur.
WARNING:
To ensure proper levels of ventilation, periodically measure the blood gas. Failure to do so could
result in patient injury.
Содержание RW82
Страница 2: ...This page intentionally left blank ...
Страница 4: ...This page intentionally left blank ...
Страница 16: ...PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04 1 8 This page intentionally left blank ...
Страница 118: ...PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04 6 10 Pneumatic System Pneumatic System Block Diagram ...
Страница 124: ...PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04 7 4 This page intentionally left blank ...
Страница 125: ...This page intentionally left blank ...