PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
2 - 6
WARNING:
The internal adjustable airway pressure relief is not functional with a self-inflating bag because
the patient supply is connected to the fresh gas reservoir on the self-inflating bag rather than
directly to the patient airway. When using a self-inflating bag, always use a suitable external
airway pressure relief valve connected to the patient airway. To obtain maximum flow, turn the
adjustable airway pressure relief knob to the
Max.
position. Failure to do so could result in
patient injury.
WARNING:
Using a breathing circuit other than the one supplied may inadvertently restrict the gas flow
such that the indication of the gas flow and the adjustable airway pressure relief are inaccurate.
Patient injury could occur.
•
Patient breathing and supply circuits
—Use the disposable breathing circuit (MU10900) for manual
bagging on the Resuscitation Module.