PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
5 - 15
WARNING:
Use only the skin probes with reflective probe covers supplied by the manufacturer. Failure to
do so could result in patient injury or equipment damage.
WARNING:
Do not place objects between the infant and the warmer module that block heat transfer and
absorb heat. The heater will directly heat the objects instead of the patient. Patient injury could
occur.
WARNING:
Locate the skin probe so that the skin around the sensor is in direct line with the heat from the
warmer module. If the location is shadowed, such as by the infant body, overheating and
possible burning of the infant skin could occur.
WARNING:
The skin probe must be in direct contact with the skin to provide accurate monitoring of the
infant skin temperature. Failure to maintain direct skin contact could result in overheating and
possible burning.
WARNING:
Check the infant condition at least every 15 min for correct sensor attachment and for signs of
overheating. Failure to do so could result in personal injury.
WARNING:
To prevent false temperature indications, place the
skin probe at least 3.8 cm (1½") from any
transcutaneous monitor (TcPO
2
or TcPCO
2
) probe,
and cover the skin probe with a reflective probe
cover. Do not place the skin probe on an area
previously used by a TcPO
2
or TcPCO
2
probe.
Patient injury could occur.
3. Attach the skin probe to the infant:
a. Before the probe is placed on the skin, thoroughly
clean and dry the skin area around the probe.
b. Place the probe on the infant abdomen, halfway
between the xyphoid and the umbilicus. If the
infant is prone, locate the skin probe on the infant
back.
c. Place the metal side of the probe in direct contact
with the skin.
Содержание RW82
Страница 2: ...This page intentionally left blank ...
Страница 4: ...This page intentionally left blank ...
Страница 16: ...PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04 1 8 This page intentionally left blank ...
Страница 118: ...PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04 6 10 Pneumatic System Pneumatic System Block Diagram ...
Страница 124: ...PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04 7 4 This page intentionally left blank ...
Страница 125: ...This page intentionally left blank ...