PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
5 - 4
Silence/Reset
key
Has three functions: silences alarms, resets Manual Mode, and
sets procedural silence.
If an alarm sounds in Manual Mode, press to silence the
High
Temp
alarm for 2 min, reset the
Chk Patient
alarm, and silence
the
Chk Patient
alarm at any time after 10 min. If it is not
pressed, the heater turns off after 15 min. Press to restore heater
power.
If an alarm sounds in Baby Mode, press to silence the
High Temp
alarm for 2 min or the
Baby Temp
alarm for 10 min. During non-
alarm conditions, press to enter procedural silence.
Pre-Warm
indicator
Indicates that the controller is operating in Pre-Warm Mode.
Manual
indicator
Indicates that the controller is operating in Manual Mode.
Baby
indicator
Indicates that the controller is operating in Baby Mode.
Mode Select
key
Press to select Pre-Warm, Manual, or Baby Mode of operation.
NOTE:
Defaults to Pre-Warm upon power up.
Skin probe connector
Accepts the skin probe for monitoring the infant skin temperature.
When connected, the
Baby Temperature
display indicates the
temperature sensed by the probe when in Manual or Baby Modes.
When in Pre-Warm Mode for software version 1.08 or greater, the
display indicates “PrE.” When the probe is disconnected, the
Baby Temperature
display goes blank in Manual or Baby Modes
and indicates "PrE" in Pre-Warm Mode. When disconnected in
Baby Mode, the probe alarm also occurs.
Keypad Lock
key
Deactivates keypad to avoid accidental changes to settings.
Press to deactivate the
>37 °C
, Up arrow, Down arrow, and
Mode
Select
keys. The
Keypad Lock
key illuminates to indicate the
keypad is locked. Press again to activate the keys.
Examination Light
key
Press to turn the examination light located in the warmer module
on or off.
Name
Description
Содержание RW82
Страница 2: ...This page intentionally left blank ...
Страница 4: ...This page intentionally left blank ...
Страница 16: ...PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04 1 8 This page intentionally left blank ...
Страница 118: ...PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04 6 10 Pneumatic System Pneumatic System Block Diagram ...
Страница 124: ...PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04 7 4 This page intentionally left blank ...
Страница 125: ...This page intentionally left blank ...