PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
2 - 27
Electromagnetic Compatibility (EMC) Guidance and Manufacturer Declarations
Guidance and Manufacturer Declaration—Electromagnetic Emissions
The Resuscitaire Radiant Warmers are intended for use in the electromagnetic environment specified
below. The customer or user of the unit should ensure that the unit is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment—Guidance
Radio frequency (RF)
emissions—CISPR 11
Group 1
The Resuscitaire Radiant Warmers use RF
energy only for its internal function. Therefore,
its RF emissions are very low and are not likely
to cause interference with nearby electronic
equipment.
RF emissions—CISPR
11
Class A
The Resuscitaire Radiant Warmers are suitable
for use in all establishments other than domestic
and those directly connected to the public low-
voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic Emissions—
IEC 61000-3-2
Not applicable
Voltage fluctuations/
flicker emissions—IEC
61000-3-3
Not applicable