1.3 AC Fail System
Switching the Unit Off
On switching the unit power off, the mains fail alarm will operate
automatically indicated by ‘AC Fail’ on the display. Simply press
the MUTE button to silence the alarm.
AC Fail System
The unit will give an AC Fail alarm whenever the mains power is removed for any reason which consists of a display ‘AC
Fail’ and an audible sound. On reconnection of the mains power the unit will automatically continue its current operation
without a full restart, as long as the power was interrupted for less than 10 seconds and the MUTE button had not been
pressed during the AC Fail alarm. This enables the system to recover from short term mains power dropouts without
intervention. A full restart with initialisation will occur whenever the AC Fail has been acknowledged with the MUTE button
pressed, or if the mains power was off for longer than 10 seconds.
1.4 Pressures, Comfort and Cycle Settings
All pressure readings in table are in mmHg.
Product Name
Therapy
Mode
Comfort Setting
Max. Inflate
Soft
Medium
Firm
Seated
QUATTRO Acute
Active
18
25
35
N/A
35
CLP
14
21
31
N/A
35
QUATTRO Plus
Active
22
32
44
50
50
CLP
20
29
41
46
50
(CLP: Continuous low pressure; alternative term for static mode)
QUATTRO 1-in-4 Mattress Cycle
Angle
Channel
Function
0º
3
Equalise
30º
4
Deflate
60º
4
Inflate
90º
4
Equalise
120º
1
Deflate
150º
1
Inflate
180º
1
Equalise
210º
2
Deflate
240º
2
Inflate
270º
2
Equalise
300º
3
Deflate
330º
3
Inflate
GREEN
DOT
CLOSED
OFF
AIR
LOSS
1
2
3
4
Содержание QUATTRO Acute
Страница 1: ...QUATTRO Acute QUATTRO Plus Service Manual DHG HE A LT HCAR E C OM ...
Страница 49: ...SERVICE MANUAL DH G HEALTHCARE COM 4 8 ...
Страница 50: ...QU ATTRO ACUTE QUATTRO PLUS DH G HEALTHCARE COM 4 9 ...
Страница 51: ...SERVICE MANUAL DH G HEALTHCARE COM 5 0 ...
Страница 53: ...SERVICE MANUAL DH G HEALTHCARE COM 5 2 4 5 Type 19R Circuit Diagrams ...
Страница 57: ...SERVICE MANUAL DH G HEALTHCARE COM 5 6 Notes ...
Страница 58: ...QU ATTRO ACUTE QUATTRO PLUS DH G HEALTHCARE COM 5 7 Notes ...
Страница 59: ...SERVICE MANUAL DH G HEALTHCARE COM 5 8 Notes ...