S E R V I C E M A N U A L
D H G - H E A LT H C A R E . C O M
2 4
1
HOLD
1. Switch the power unit on with information (data) key depressed.
The unit will display ‘QUATTRO PLUS or ACUTE Information’.
2. Press the down arrow key to scroll through the information menu
until the unit displays ‘LAST SERV. DATE’ and the current date.
3. Press the active/CLP toggle select key and keep it pressed for
5 seconds until a row of stars are displayed where the date was
displayed.
4. Once the stars have been displayed press the information (data) key
and the unit will display the date with a cursor.
5. Use the active/CLP toggle select key to move the cursor to the
parts of the date that need altering, and the down and up arrow
keys
to alter the date itself. Once the date is correct press
information (data) key to exit the menu and the cursor will no
longer display.
3
Altering Service Date
SCROLL
2
HOLD
4
5
2.8 Uploading New Software with Service Data USB V1.2
1. Connect the unit to power but do not switch on and connect the data lead coming from USB communication
box to the unit and communication box to the computer as shown in pictures below. (LEAD COLOURS MIGHT BE
DIFFERENT).
2. Start-up the service data program,
and on service data and select the
“upload program” tab shown.
2
1
Data
lead
Date lead
USB
cable
OR
Содержание QUATTRO Acute
Страница 1: ...QUATTRO Acute QUATTRO Plus Service Manual DHG HE A LT HCAR E C OM ...
Страница 49: ...SERVICE MANUAL DH G HEALTHCARE COM 4 8 ...
Страница 50: ...QU ATTRO ACUTE QUATTRO PLUS DH G HEALTHCARE COM 4 9 ...
Страница 51: ...SERVICE MANUAL DH G HEALTHCARE COM 5 0 ...
Страница 53: ...SERVICE MANUAL DH G HEALTHCARE COM 5 2 4 5 Type 19R Circuit Diagrams ...
Страница 57: ...SERVICE MANUAL DH G HEALTHCARE COM 5 6 Notes ...
Страница 58: ...QU ATTRO ACUTE QUATTRO PLUS DH G HEALTHCARE COM 5 7 Notes ...
Страница 59: ...SERVICE MANUAL DH G HEALTHCARE COM 5 8 Notes ...