S E R V I C E M A N U A L
D H G - H E A LT H C A R E . C O M
5 4
Guidance and Manufacturer’s Declaration: Electromagnetic Immunity (IEC 60601-1-2)
Immunity Test
IEC 60601 Test
Level
Compliance Level
Electromagnetic
Environment - Guidance
Electrostatic discharge (ESD) IEC
61000-4-2
± 8kV contact
± 15kV air
± 8kV contact
± 15kV air
The relative humidity should be
at least 5%.
Electrical fast transient/burst IEC
61000-4-4
± 2 kV For mains
supply lines
100kHz repetition
frequency
± 2 kV For mains supply
lines
100 kHz repetition
frequency
Mains supply quality for the
mains adapter should be that
of a typical commercial and/or
hospital environment.
Surge
IEC61000-4-5
± 2kV Line(s) to
ground
± 1kV line(s) to line
± 2kV Line(s) to ground
± 1kV line(s) to line
Mains supply quality for the
mains adapter should be that
of a typical commercial and/or
hospital environment.
Voltage dips, short interruptions
and voltage variations on mains
supply IEC 61000-4-11
0 % U
T
; 0,5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % U
T
; 0,5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
Mains power quality should be
that of a typical commercial
and/or hospital environment.
If the user of the QUATTRO
requires continued operation
during power mains interruption,
it is recommended that the
QUATTRO be powered from an
uninterruptible power supply or
battery.
0 % U
T
; 1 cycle
70 % U
T
; 25/30
cycles
Single phase: at 0°
0 % U
T
; 1 cycle
70 % U
T
; 25/30 cycles
Single phase: at 0°
Voltage interruptions
0 % U
T
; 250/300
cycle
0 % U
T
; 250/300 cycle
Mains frequency (50/60Hz)
magnetic field IEC61000-4-8
30 A/m
30 A/m
Mains frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial and/or hospital
environment.
Note: U
T
is the A.C. mains voltage prior to application of the test level.
Guidance and Manufacturer’s Declaration: Electromagnetic Emissions (IEC 60601-1-2)
Emissions Test
Compliance
Electromagnetic Environment - Guidance
RF emissions
CISPR 11
Class B
The QUATTRO systems are suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low-
voltage power supply network that supplies buildings
used for domestic purposes.
Harmonics emissions
61000-3-2
Class A
Voltage fluctuations / flicker emissions
61000-3-3
Complies
Declaration – Electromagnetic Emissions
Declaration – Electromagnetic Immunity
Содержание QUATTRO Acute
Страница 1: ...QUATTRO Acute QUATTRO Plus Service Manual DHG HE A LT HCAR E C OM ...
Страница 49: ...SERVICE MANUAL DH G HEALTHCARE COM 4 8 ...
Страница 50: ...QU ATTRO ACUTE QUATTRO PLUS DH G HEALTHCARE COM 4 9 ...
Страница 51: ...SERVICE MANUAL DH G HEALTHCARE COM 5 0 ...
Страница 53: ...SERVICE MANUAL DH G HEALTHCARE COM 5 2 4 5 Type 19R Circuit Diagrams ...
Страница 57: ...SERVICE MANUAL DH G HEALTHCARE COM 5 6 Notes ...
Страница 58: ...QU ATTRO ACUTE QUATTRO PLUS DH G HEALTHCARE COM 5 7 Notes ...
Страница 59: ...SERVICE MANUAL DH G HEALTHCARE COM 5 8 Notes ...