2.2 Test Menu Options and Navigation
1. Apply the power lead to unit and switch the unit on, with the down
arrow key depressed at the same time.
Once the unit has started, release the down arrow key . This
enables you to enter the test menu.
2. The unit will display ‘TEST MODE CAL’. Test mode calibration is to
recalibrate pressure sensor circuit on PCB.
Use either the up or down arrow keys to scroll through the
test menu. For the purpose of this document the down arrow key was
used.
‘TEST MODE Test LCD’
This test enables you to test functionality of all the segments of LCD and
the unit’s sounder.
‘TEST MODE Test Keys’
This test enables you to test the functionality of all six keys. Enter “Test
Keys” and press each key to check functionality.
‘TEST MODE Test Pump’
This test mode enables you to test functionality of pump.
‘TEST MODE Rotor System’
This test enables you to check functionality of CCV assembly.
‘TEST MODE Test PReg’
This test enables you to check the pressure regulation functionality of
your QUATTRO.
‘TEST MODE Battery Volts’
This test enables you to check that the on board battery is functioning
correctly.
Enter ‘Battery Volts test’ and the unit will display battery volts value this
voltage value must be rising showing that the battery is charging.
‘TEST MODE Set Language’
This enables you to select the language that the unit displays.
S E R V I C E M A N U A L
D H G - H E A LT H C A R E . C O M
1 6
1
HOLD
2
OR
Содержание QUATTRO Acute
Страница 1: ...QUATTRO Acute QUATTRO Plus Service Manual DHG HE A LT HCAR E C OM ...
Страница 49: ...SERVICE MANUAL DH G HEALTHCARE COM 4 8 ...
Страница 50: ...QU ATTRO ACUTE QUATTRO PLUS DH G HEALTHCARE COM 4 9 ...
Страница 51: ...SERVICE MANUAL DH G HEALTHCARE COM 5 0 ...
Страница 53: ...SERVICE MANUAL DH G HEALTHCARE COM 5 2 4 5 Type 19R Circuit Diagrams ...
Страница 57: ...SERVICE MANUAL DH G HEALTHCARE COM 5 6 Notes ...
Страница 58: ...QU ATTRO ACUTE QUATTRO PLUS DH G HEALTHCARE COM 5 7 Notes ...
Страница 59: ...SERVICE MANUAL DH G HEALTHCARE COM 5 8 Notes ...