Q U AT T R O
®
A C U T E & Q U AT T R O
®
P L U S
D H G - H E A LT H C A R E . C O M
4 1
3.12 Mattress Umbilical Replacement
1. Remove the cover, deflate the mattress, and unfasten the cells
in the proximity of the umbilical (see section 3.10).
2. Identify the tubes that go to the umbilical for ease when
reassembling, then remove the tubes from the tube mounts
and fully unscrew the M8 cap head alun
screw in the centre of the flange .
3.
Remove the inner umbilical plate and then
remove the umbilical. Apply the reverse
procedure for umbilical replacement.
4. When reassembling, the new umbilical
tubes should be attached in the
orientation shown.
UMBILICAL WITH
BARB AND SLEEVE
ASSEMBLY QA
G4 SPARE: PART
NUMBER 12198
UMBILICAL WITH
BARB AND SLEEVE
ASSEMBLY QP
G4 SPARE: PART
NUMBER 12197
A
1
4
2
3
A
2
4
1
AL
3
1. Deflate the mattress, remove the mattress cover and unfasten
the end and centre poppers of all the air cells (see section 3.10).
2. Remove the CPR as shown in section 3.11.
3. Remove the umbilical as shown in section 3.12.
4.
Remove the air cell set and tubes from the base sheet and apply
the reverse procedure for mattress reassembly.
SEE SPARES LIST FOR APPROPRIATE BASE SHEET REQUIRED
FOR MATTRESS.
3
2
3.13 Mattress Base Sheet Replacement
1
4
Содержание QUATTRO Acute
Страница 1: ...QUATTRO Acute QUATTRO Plus Service Manual DHG HE A LT HCAR E C OM ...
Страница 49: ...SERVICE MANUAL DH G HEALTHCARE COM 4 8 ...
Страница 50: ...QU ATTRO ACUTE QUATTRO PLUS DH G HEALTHCARE COM 4 9 ...
Страница 51: ...SERVICE MANUAL DH G HEALTHCARE COM 5 0 ...
Страница 53: ...SERVICE MANUAL DH G HEALTHCARE COM 5 2 4 5 Type 19R Circuit Diagrams ...
Страница 57: ...SERVICE MANUAL DH G HEALTHCARE COM 5 6 Notes ...
Страница 58: ...QU ATTRO ACUTE QUATTRO PLUS DH G HEALTHCARE COM 5 7 Notes ...
Страница 59: ...SERVICE MANUAL DH G HEALTHCARE COM 5 8 Notes ...