Q U AT T R O
®
A C U T E & Q U AT T R O
®
P L U S
D H G - H E A LT H C A R E . C O M
Product
Key 1
Key 2
Key 3
Key 4
Key 5
Key 6
QUATTRO Acute
Static (CLP)
/ dynamic
toggle
Decrease
comfort
setting
Increase
comfort
setting
Information
Mute /
Unlock
Transport
mode (max
inflate)
QUATTRO Plus
Static (CLP)
/ dynamic
toggle
Decrease
comfort
setting
Increase
comfort
setting
Information
Mute /
Unlock
Transport
mode (max
inflate)
THERAPY MODE (Key 1 - Static (CLP) / dynamic toggle)
Pressing the THERAPY MODE button will toggle between ACTIVE (1 in 4 alternating air pressure cycle) and
CONTINUOUS LOW PRESSURE therapy modes. The selected therapy mode is shown on the display screen.
The default mode is ACTIVE therapy. To switch to CONTINUOUS LOW PRESSURE mode (once unlocked*), press THERAPY
MODE button for 2 seconds until an audible tone is heard. The system will then display ‘PLEASE WAIT’ and requires
approximately 2 minutes to initialise.
COMFORT CONTROL (Key 2 - Decrease comfort setting / Key 3 - Increase comfort setting)
Air pressure is regulated within each of the cells throughout the cycle so that support, posture and therapy are
constantly maintained at optimum levels, in response to patient weight, movement and position. Equalisation
of cell pressure automatically takes place at each stage of the alternating air pressure cycle, again to ensure
precise pressure and therapy is provided. The automatic default comfort setting is MEDIUM. However, if the
patient prefers a firmer or softer mattress, increase or decrease the comfort control setting accordingly (once
unlocked*) using the UP and DOWN arrow buttons (SOFT/MEDIUM/FIRM). The comfort setting is shown on the display
screen. Check periodically to ensure patient support and comfort.
DATA (Key 4 - Information) (Used for accessing information only, does not affect mode of operation)
Pressing the DATA button at any time switches the display into DATA mode. Use the up and down arrow buttons
to scroll through the product data and user information set. Pressing the DATA button again returns the display
to the previous mode.
*MUTE/UNLOCK (Key 5)
Press to silence the sounder and to clear the message from the display screen. The power unit will automatically
lock 2 minutes after the last button operation when running to prevent the inadvertent operation of button
functions (except MUTE), as indicated by ‘ ‘ on the display screen. Press and hold the MUTE/UNLOCK button until the
power unit beeps if further button operation is needed (i.e. comfort setting). The power unit will lock again 2 minutes
after the last button operation.
NB. After power failure/switching the power off, pressing MUTE cancels the system’s previous settings. When power
returns the default setting of ACTIVE mode, MEDIUM comfort setting is invoked. (Note that previous settings are
automatically cancelled if the duration between switch off and switch on is greater than 12 seconds. If power returns
before a period of 12 seconds has passed and the MUTE button has not been pressed, the system will return to the
previous mode of operation.)
MAX. INFLATE (Key 6 - Transport mode / max inflate)
Necessary for some nursing procedures, the MAX INFLATE mode inflates the mattress to maximum static
pressure for a period of 15 minutes. After pressing the MAX INFLATE button (once unlocked*) to inflate
mattress, the system displays ‘PLEASE WAIT’ followed by ‘READY’ and a 5 second audible tone when maximum pressure
is achieved and ‘MAX INFLATE’ is shown on the display screen. After 15 minutes the system automatically returns to the
ACTIVE mode of operation.
5
(CLP: Continuous low pressure; alternative term for static mode)
Содержание QUATTRO Acute
Страница 1: ...QUATTRO Acute QUATTRO Plus Service Manual DHG HE A LT HCAR E C OM ...
Страница 49: ...SERVICE MANUAL DH G HEALTHCARE COM 4 8 ...
Страница 50: ...QU ATTRO ACUTE QUATTRO PLUS DH G HEALTHCARE COM 4 9 ...
Страница 51: ...SERVICE MANUAL DH G HEALTHCARE COM 5 0 ...
Страница 53: ...SERVICE MANUAL DH G HEALTHCARE COM 5 2 4 5 Type 19R Circuit Diagrams ...
Страница 57: ...SERVICE MANUAL DH G HEALTHCARE COM 5 6 Notes ...
Страница 58: ...QU ATTRO ACUTE QUATTRO PLUS DH G HEALTHCARE COM 5 7 Notes ...
Страница 59: ...SERVICE MANUAL DH G HEALTHCARE COM 5 8 Notes ...