3.5 Cycle Control Valve (CCV) Replacement
1. Remove the rear case assembly via the 10 pozi-drive screws
(see section 3.2).
2. Withdraw the double pump unit, disconnecting the air hose to
the cycle control valve
and the power connector on the
PCB .
3. Disconnect the clear sensor tube from the CCV (not the PCB).
4. Disconnect the crimp housing for the mains power from
the control PCB header (this will pull away from the PCB).
Disconnect the ribbon cable - note this is a locking 4 way
header, press in on the locking mechanism release tab and
pull away from the PCB.
5. The CCV assembly can now be lifted from the enclosure and
replaced with new T19R CCV assembly.
6. When replacing the pump or cycle control valve assembly it is
important to ensure they are reconnected to the correct header.
Both headers are marked on PCB “PUMP” and “MOTOR”.
1
2
B
A
CCV
ASSEMBLY
230:50
T19R: PART
NUMBER
10944
A
B
3
4
C
C
D
Push on
release
tab and
lift
D
5
6
Содержание QUATTRO Acute
Страница 1: ...QUATTRO Acute QUATTRO Plus Service Manual DHG HE A LT HCAR E C OM ...
Страница 49: ...SERVICE MANUAL DH G HEALTHCARE COM 4 8 ...
Страница 50: ...QU ATTRO ACUTE QUATTRO PLUS DH G HEALTHCARE COM 4 9 ...
Страница 51: ...SERVICE MANUAL DH G HEALTHCARE COM 5 0 ...
Страница 53: ...SERVICE MANUAL DH G HEALTHCARE COM 5 2 4 5 Type 19R Circuit Diagrams ...
Страница 57: ...SERVICE MANUAL DH G HEALTHCARE COM 5 6 Notes ...
Страница 58: ...QU ATTRO ACUTE QUATTRO PLUS DH G HEALTHCARE COM 5 7 Notes ...
Страница 59: ...SERVICE MANUAL DH G HEALTHCARE COM 5 8 Notes ...